Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
453
System configuration
To configure the function keys
In the following steps, the letters in parentheses
correspond to the diagram for the
Config. buttons
page (see page 452
).
1
Select the
System setup...
button on the main
menu bar.
2
Select the
Screen setup
tab (if not already
selected).
3
Select the
Config. buttons
tab. A password
popup appears.
4
Enter the clinical password and select the
Enter
button.
5
Select the button on the menu bar list to be
replaced with another one.
The button to be removed is highlighted in
yellow. A pop
-
up window (P) with a slide ruler
appears. The popup contains the following list
of available selections. Up to 14 buttons can be
assigned to the main menu bar:
NOTE
Any button appearing on a light gray background
identifies one that holds a permanent position on
the menu bar. These buttons cannot be ex-
changed and are therefore not selectable.
Button
Description
None
– Reassigns the button name to’
None
’ in the
Config. buttons
page.
– Removes the original button from the main menu bar and reorders the
buttons on the main menu bar.
All alarms paused
Pauses all alarms at the Cockpit for two minutes.
Auto set all
Adjusts the alarm limits of all parameters automatically.
Code
Executes pre-configured functions during an emergency.
Night time
Controls the brightness of the Cockpit screen.
Default view
Resets the Cockpit to the factory-default view.
Discharge
Discharges a patient from the Cockpit.
Freeze waveforms
Stops the waveform from scrolling for approximately 60 seconds.
Home
Returns to the main screen and closes any dialog.
Mark event
Stores 20 seconds of waveform and parameter data in the alarm history.
NIBP start/stop
Starts or stops an non-invasive blood pressure measurement.
NIBP continuous
Starts or stops a continuous non-invasive blood pressure measurement.
Pacer detection
Allows you to activate or deactivate pacer detection.
Print case summary
Prints a combination of reports configured in the
Reports
page.
Print screen
Prints the contents of the current screen on a connected laser printer.
Privacy
Activates privacy mode (patient monitoring continues but the patient data
are removed from the Cockpit and the M540 and appear only at the ICS).
Relearn ARR
Prompts the M540 to learn the dominant QRS pattern of a patient to iden-
tify the rhythms as either normal or irregular.
Relearn ST
Prompts the M540 to learn the dominant ST-segment deviations of a pa-
tient in order to identify ST rhythms as either normal or irregular.