Technical data
596
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Voltage dips and
short interruptions
on AC mains input
lines
(IEC 61000-4-11)
Dip >95%, 0.5 periods
Dip 60%, 5 periods
Dip 30%, 25 periods
Dip >95%, 5 seconds
>95%,
0.5 periods
60%, 5 periods
30%, 25 periods
>95%,
5 seconds
Mains power should be a typical
commercial or hospital environment.
If user requires continued operation
during power mains interruptions,
ensure batteries are installed and
charged. Ensure battery life exceeds
longest anticipated power supply
failures or provide additional
uninterruptible power supply.
Electromagnetic immunity
Immunity
against...
IEC 60601-1-2 test level Compliance
level (of device)
Electromagnetic environment