Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
123
Alarms
Activating or deactivating acoustic alarm signals
Deactivating acoustic alarm signals is a password-
protected function (see page 458).
You can deactivate alarm tones permanently. When
you deactivate alarm tones, the following happens:
– Alarm tones no longer announce alarm
conditions.
– The message
Audio off
appears in the far right
field of the header bar and the following symbol:
After deactivating acoustic alarm signals
permanently, you can activate them again (see
page 458). When you activate acoustic alarm
signals, the following happens when an alarm
condition occurs:
– Acoustic alarm signals sound (see page 114).
– Alarm messages appear in the header bar (see
page 112).
For information about configuring alarm volumes
and the
Audio off reminder
feature, refer to "Alarm
setup – configuring the alarm volume and tones"
on page 461.
Pausing alarm monitoring temporarily
If the password-protected alarm pause feature is
activated (see page 459), you can pause alarm
monitoring temporarily. The alarm pause duration is
adjustable from 1 minute to 5 minutes (default is
2 minutes).
The following happens when you pause alarm
monitoring:
– Acoustic and optical alarm signals for new
alarm conditions are suppressed for all
parameters until alarm monitoring begins again.
– Alarm signals for any active alarm condition
stop immediately.
– The alarming parameter field and alarm bar
return to the pre-alarm state.
– Alarm messages are removed from the alarm
message field in the header bar.
– The far right field of the header bar turns yellow
and displays the alarm message
All alarms
paused
, a timer, and the following symbol:
– The message
All alarms paused
is recorded in
the alarm history (see page 139).
NOTE
If the
French NFC mode
is activated (see
page 473), you cannot pause alarm monitoring for
more than 3 minutes.
NOTE
If the Cockpit is connected to the network and the
patient is admitted at the ICS, a message also
appears at the ICS that alarms are paused.