Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
389
External Device – Neuromuscular transmission (NMT)
NMT parameter field (TOF/TOFS mode)
The NMT parameter field in ’TOF /TOFs’ mode is
very similar and contains the following elements:
A
Neuromuscular transmission label (
NMT
)
B
NMT parameter label
TOF Ratio
or
TOF Cnt
TOF Ratio
appears when the TOF count is
equal to four twitches and the amplitude of the
first twitch is
≥
20%.
TOF Cnt
appears when the TOF count is
≤
than
three twitches or there are four twitches and the
magnitude of the first twitch is < 20%.
C
Time stamp of the current value (the time stamp
identifies the time the Cockpit received the
value from the connected device)
D
Count down time bar and value indicating the
remaining time in the interval before the start of
the next set of measurements - displayed only
in TOFs mode
The label
auto
15 s
appears instead of the
count down bar when the user initiates
automatic TOF.
E
TOF Ratio / TOF-Cnt value
F
4 individual amplitude bar graphs indicating the
number of TOF counts (the last bar shows the
magnitude of the fourth twitch).
56
3
A
B
C
30
D
E
F