![Dräger Infinity Acute Care System M540 Скачать руководство пользователя страница 513](http://html.mh-extra.com/html/drager/infinity-acute-care-system-m540/infinity-acute-care-system-m540_instructions-for-use-manual_2526773513.webp)
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
513
Troubleshooting
Messages
Priority
Message
Cause
Remedy
!!!
All alarms off
The
All alarms paused
function is
set to
No timeout
(see page 458)
and the
All alarms off
button was
selected.
Select the
All alarms off
button again to remove the
message.
!!!
All alarms paused
with timer
The
All alarms paused
function is
set to a time (see page 458) and the
All alarms paused
button was
selected.
Select the
All alarms
paused
button again to
remove the message.
!!!
ASY, VF off
This message appears in the alarm
message field under the following
circumstances:
– Heart rate alarms are enabled
–
HR source
is set to
ART
or
SpO2
–
ARR mode
(arrhythmia) is set
to
Off
The message disappears
under the following
circumstances:
– The setting
HR source
is changed to
ECG
– The setting
ARR mode
is changed to
Basic
or
Advanced
.
!!
Audio paused
with
timer
The yellow
Audio paused 2 min
key (next to the rotary knob) was
pressed.
Press the key again to
remove the message.
!!
Audio off
This message appears in the alarm
message field when the
Audio off
feature is set to
On
.
Activate the
Audio off
setting to remove the
message.
!!!
All alarms off:
bypass
This message appears in the alarm
message field when you activate
cardiac bypass mode (see
page 460).
Deactivate the feature to
remove the message.
None
Discharge
Touch Screen to
resume monitoring
This message appears in the center
of the Cockpit screen when the
patient has been discharged (see
page 95).
Touch the screen to resume
monitoring and admit a new
patient.
!
Duplicate IP address
This message appears in the alarm
message field when a duplicate IP
address is detected anywhere on
the Infinity network.
The Cockpit goes offline within 10
seconds of a
Duplicate IP address
alarm condition.
Configure a new IP
address.
The Cockpit then
immediately tries to rejoin
the Infinity network