ECG, arrhythmia, and ST segment
220
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
The arrhythmia monitoring functions have configu-
rable parameter-specific setup pages (see
page 225).
Before performing any monitoring functions, refer
to the section “For your safety and that of your
patients“on page 13.
Selecting arrhythmia leads
Appropriate lead selection is essential for accurate
arrhythmia monitoring. Ideally, the two best leads
should be assigned to the top two waveform
channels.
The following two selections are available:
–
ECG1
(
single channel selection)
–
dedicates
processing to the lead in the top channel.
–
ECG1 & 2
(
dual channel selection)
–
determines the heart rate and ARR based on
the leads in the two top channels.
To select arrhythmia leads
Select the heart rate parameter field to select
the
ECG
page directly.
or
1
Select
Sensor parameters...
from the main
menu bar.
2
Select the
ECG
tab to access the
ECG
page.
3
Select the
Settings 2
tab.
4
Select the button next to
ECG1
or
ECG1 & 2
and select the desired lead.
WARNING
The message
HR alarms off
appears in the
right most field in the header bar whenever
you deactivate heart rate alarms.
The
HR, ASY, VF off
message appears
when arrhythmia monitoring is deactivated,
the
ASY/VF alarms
feature is set to
Follow HR
alarm
, and heart rate alarms are deactivated.
NOTE
If French NFC mode is activated (see page 473),
you cannot deactivate heart rate alarms.