Temperature
284
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Connecting the temperature sensors to the M540
You can connect a single sensor or two sensors to
the M540 directly using the dual temperature Y-
cable. The dual temperature Y-cable monitors two
temperatures simultaneously.
A
M540 temperature port
B
Dual temperature Y-cable
C
Dual temperature Y-cable connectors
D
Single temperature sensor
E
Temperature sensor connecting directly to the
M540
To connect two temperature sensors
1
Connect the temperature sensors (D) to the
connectors (C) of the dual temperature Y-cable.
2
Connect the connector (B) of the dual
temperature Y-cable to the M540 temperature
port (A).
To connect a single temperature sensor
Connect a temperature sensor (E) directly to
the M540 temperature port (A).
333
A
B
C
C
D
E