Alarms
134
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Configuring the arrhythmia alarm setup
The following diagram shows the
Limits
>
ARR
page for configuring the alarm settings for
arrhythmia parameters. This page consists of a
table with setup rows for each arrhythmia
parameter. Each setup row consists of several
fields for configuring the individual ARR alarm
settings. When you select a field on this page, an
orange frame highlights the selected row.
A
Limits
tab
B
Arrhythmia
category column for identifying the
ARR label
C
Alarm
priority column for selecting an alarm pri-
ority
D
Rate
column for setting the rate
E
Count
column for setting the count
F
Archive
column
G
ARR
tab
H
Arrhythmia mode
buttons
I
Relearn
button
Configuring ARR alarm settings
In addition to ARR alarm settings, the
Limits
>
ARR
page also allows you to select the arrhythmia
mode (see page 221) and initiate the relearn
process of ECG leads (see page 236). In the
following steps, the letters in parentheses refer to
the diagram of the
Limits
>
ARR
page. Alarm
ranges and defaults are listed starting on page 160.
01
9
Alarms
A
B
C
D
E
H
H
F
G
H
I