![Dräger Infinity Acute Care System M540 Скачать руководство пользователя страница 252](http://html.mh-extra.com/html/drager/infinity-acute-care-system-m540/infinity-acute-care-system-m540_instructions-for-use-manual_2526773252.webp)
SpO2 and Pulse CO-Ox monitoring with Masimo SET MCable
252
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Overview of SpO2
and
Pulse CO-Ox
monitoring
SpO2 and Pulse CO-Ox monitoring is only possible
with the corresponding MCable. The following
hardware is available from Masimo for monitoring
SpO2 and
Pulse CO-Ox
parameters.
– Infinity MCable – Masimo SET (Masimo SET
MCable)
– Infinity MCable – Masimo rainbow SET
(Masimo rainbow SET MCable)
The values and the waveform are displayed on the
M540 and on the Cockpit.
The Masimo SET MCable and Masimo rainbow
SET MCable support motion tolerant pulse
oximetry using Signal Extraction Technology (SET).
This technology enhances the quality of SpO2
monitoring and also measures the percentage of
functional hemoglobin saturated with oxygen
(%SpO2) in the arterial blood of the patient
accurately and effectively.
A sensor applied to the patient measures the
absorption levels of red and infrared light. The
Masimo SET MCable or Masimo rainbow SET
MCable uses the difference between the two
measurements to calculate the percentage of
saturated hemoglobin (SpO2). Because light
absorption varies with blood volume and blood
volume varies with pulse rate, both types of
Masimo SET MCable can also derive a pulse rate
(PLS).
In addition, the Masimo SET MCable also provides
a perfusion index (PI) value. PI is the ratio of the
pulsatile blood flow to the non-pulsatile blood flow
in peripheral tissue. The PI value provides
information regarding the perfusion status of the
selected application site. This provides a means to
select the most optimal site.
The Infinity MCable – Masimo rainbow SET
measures additional parameters that continuously
and non-invasively measure blood constituents.
SpO2 and
Pulse CO-Ox
measurements are for
adult, pediatric, and neonatal patients (with the
following exceptions).
Refer to the instructions for use
Infinity Acute Care
System – Infinity M540
for a detailed description of
the M540 SpO2 functions. The SpO2 monitoring
functions are configurable in the parameter-specific
setup page (see page 264).
Before performing any monitoring functions, refer
to the section "For your safety and that of your
patients" on page 13. Parameter-specific error
messages are listed on page 526.
Supported parameters
The parameters SpO2, PLS, and PI are available
and displayed regardless of which Masimo sensor
and which Masimo SET MCable is being used.
The availability of additional Masimo rainbow SET
parameters depends on the sensor type that is
connected and which parameters are activated on
the Masimo rainbow SET MCable.
NOTE
The Masimo rainbow SET MCable parameters
SpHb and SpOC are not approved for neonatal
monitoring.
NOTE
Information about wavelength range may be
useful during photodynamic therapy. For details,
see the technical data chapter of the instructions
for use
Infinity Acute Care System – Infinity M540
.
NOTE
This device is covered under one or more of the
following USA patents: 5,758,644, 6,011,986,
6,699,194, 7,214,986, 7,254,433, 7,530,955 and
other applicable patents listed at:
www.masimo.com/patents.htm