System configuration
460
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
SpO2 alarm delay
–
On
(default)
– Off
The alarm validation feature must be activated
to use this setting.
When this setting is activated, an SpO2 lower
alarm limit violation must persist for 10 seconds
before triggering acoustic and optical alarm
signals.
This function is not possible if the Nellcor
SatSeconds alarm
feature
is set to any value
other than
Off
(see page 279).
Show alarm limits
–
On
(default)
– Off
Determines whether alarm limits appear in the
parameter fields.
Alarm bar enabled
–
On
(default),
– Off
Determines whether the alarm bar flashes
during an alarm.
OR Alarms
–
On
,
–
Off
(default)
Activates/deactivates OR alarms. Alarm
functions are affected when OR alarms is
activated (see page 119).
Cardiac bypass
– On
–
Off
(default)
Activates/deactivates cardiac bypass mode.
Alarm functions are affected when cardiac
bypass mode is activated (see page 119).
This mode is not available when the
French
NFC mode
is enabled (see page 473).
NIBP/SpO2
interlock
– On
–
Off
(default)
On
– the SpO2 alarm function is deactivated
during non-invasive blood pressure and Pulse
CO-Ox measurements (for more details, see
“NIBP/SpO
2
interlock alarms“on page 117).
Off
– the SpO2 alarm function is activated
during NIBP and Pulse CO-Ox measurements.
ASY/VF alarms
– Always on
(default)
– Follow HR alarm
Always on
– the ASY/VF alarm functions are
always activated.
Follow HR alarm
– the ASY and VF alarm
settings follow the setting of the heart rate
alarms.
Pacer detection
mode
– Advanced
–
Basic
(default)
Advanced
– you can select fusion mode in the
ECG
page (see page 214).
Basic
– fusion mode is not selectable.
Selection
Available settings