Troubleshooting
536
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
!
%0 cuff leak
1)
The message
Last
measurement
failed!
is followed
by the message
Cuff leak
Parameter values
are replaced by ***
The drop in cuff
pressure at the
end of the
inflation cycle is
too great.
– Check the hose
and cuff for leaks.
Replace if
necessary.
– Restart the
measurement. If
the message does
not clear, contact
DrägerService
®
.
!
%0 measurement
timeout
1)
Parameter values
are replaced by ***
An NIBP
measurement
has exceeded
time-out limit.
Repeat the
measurement.
!
%0 overpressure
1)
Parameter values
are replaced by ***
The cuff pressure
has exceeded
the overpressure
threshold.
– Check the patient
and treat if
necessary.
– Check the cuff for
obstructions.
– Repeat the
measurement.
1)
%0 is a placeholder for the parameter label NIBP.
!
%0 open line
1)
The message
Last
measurement
failed!
is followed
by the message
Open line
Parameter values
are replaced by ***
There was no
significant
increase in cuff
pressure during
the inflation
cycle.
Make sure that the
hose and cuff are
properly connected to
the monitor.
None
Venous stasis
started
Parameter value or
blank value
Message
reporting the
status of venous
stasis.
Informational message
– no action required.
None
Venous stasis
ending
Parameter value or
blank value
Message
reporting the
status of venous
stasis.
Informational message
– no action required
None
Venous stasis
ended
Parameter value or
blank value
Message
reporting the
status of venous
stasis.
Informational message
– no action required
1)
%0 is a placeholder for the parameter label NIBP.
Priority
Message
Parameter field
Problem
Solution