Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
323
Cardiac output (C.O.)
Connecting the cardiac output hardware
You can connect the hemodynamic cable to one of
the following devices:
– MPod – QuadHemo
– Hemo4 pod
– Hemo2 pod
The intermediate cable from the listed devices
connect directly to the M540.
A
M540 hemodynamic port
B
Gray connector of the hemodynamic cable
C
Red connector of the hemodynamic cable
D
MPod – QuadHemo hemodynamic port
E
Pod connector of the cardiac-output
intermediate cable
F
Cardiac output port of the MPod – QuadHemo
G
Thermistor port of the cardiac-output
intermediate cable
H
Catheter cable and thermistor cable
To connect the cardiac output hardware using
the MPod – QuadHemo
1
Connect the gray connector of the
hemodynamic cable (B) to the gray
hemodynamic port (A) of the M540.
2
Connect the red connector of the hemodynamic
cable (C) to the MPod – QuadHemo
hemodynamic port (D).
3
Connect the pod connector of the cardiac-
output intermediate cable (E) to the cardiac
output port of the MPod – QuadHemo (F).
4
Connect the catheter and the thermistor cables
(H) to the thermistor port of the cardiac-output
intermediate cable (G).
34
5
B
C
D
A
E
F
G
H