Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
311
Invasive blood pressure (IBP)
Standard labels
The M540 detects the labels automatically from the
hemodynamic pod, provided a transducer is
connected. The M540 transfers the labels to the
Cockpit. You can also label pressure channels
manually.
The following table lists the available invasive blood
pressure labels.
IBP Labels
Label
Pressure type
Measured pressures
Measurement range
ART
Arterial blood pressure
Systolic, diastolic, mean
–50 to +400 mmHg
–6.6 to +53.3 kPa
LV
Left ventricular blood
pressure
PA
Pulmonary arterial blood
pressure
RV
Right ventricular blood
pressure
CVP
Central venous blood
pressure
Mean
CPP
1)
Cerebral perfusion pressure
RA
Right atrial blood pressure
LA
Left atrial blood pressure
ICP
Intracranial pressure
Generic Labels
GP1 to GP4
Systolic, diastolic, mean
1)
The CPP value is only calculated when ICP and ART M pressure values are available.