System configuration
462
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Attention tone
volume
– Off
– 5, 10 (default) to 100% (in
increments of 10%)
Determines the volume of the attention tone or
deactivates the attention tone.
"Audio off"
reminder
–
On
(default)
– Off
Sounds an alarm tone every 30 seconds at the
Cockpit and at the M540 to remind you that the
alarm tone is deactivated during an active alarm
condition. This feature is only active when an
alarm condition exists. This alarm tone is
suppressed if you initiate an audio pause.
When the Cockpit is in OR mode, the volume of
the alarm tone corresponds to the
Alarm
volume
setting of 10%. When OR mode is not
activated, the volume equals to 50%.
This feature is not supported on remote
devices.
–
On
– a truncated acoustic alarm signal
sounds every 30 seconds for an alarm
condition of medium or high priority. Low-
priority alarms tones are not truncated.
During multiple alarm conditions, the
reminder tone adjusts itself to always report
the alarm condition with the highest alarm
priority.
–
Off
– No alarm tone sounds when the alarm
volume is deactivated and an alarm occurs.
Selection
Available settings
Description