Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
319
Invasive blood pressure (IBP)
Accessing the invasive blood pressure settings
Select the IBP parameter field to select the
IBP
page directly.
or
Select
Sensor parameters...
from the main
menu bar >
IBP
tab to access the
IBP
page. If
you cannot see the tab, select the following two
symbols located in the upper right corner of the
dialog:
symbol and the display filter
button.
The top portion of the page contains the
Auto set
and
Alarm
buttons for configuring the alarm
functions. For detailed alarm setup information, see
"Configuring the alarm settings for a patient" on
page 125.
Invasive blood pressure parameter setup functions
All invasive blood pressure setup functions take
place in the
IBP
page (see page 319).
Selection
Available settings
Description
Zero
None
Zeroes only the pressure indicated on
the
IBP
page and displays the time and
date of the last zeroing (see page 313).
Label
ART
,
PA
,
CVP
,
LA
,
LV
,
RV
,
RA
,
ICP
,
GP1
to
GP4
.
The defaults are as follows:
– Channel 1: GP1
– Channel 2: GP2
– Channel 3: GP3
– Channel 4: GP4
Allows you to assign a label to each
pressure channel.
Scale
– 5, 10, 15, 20, 25, 30, 35, 40, 50, 75, 100,
125, 150, 175, 200, 225, 250, 300 mmHg
– 1, 2, 3, 4, 5, 6, 8, 10, 12, 14, 16, 18, 20,
24, 30, 32, 36, 40 kPa
– GP1 to 4, ART, LV:
200 mmHg (30 kPa) for adults
150 mmHg (20 kPa) for pediatrics
100 mmHg (16 kPa) for neonates
– PA, RV: 50 mmHg (12 kPa)
– ICP, CVP, LA, RA: 20 mmHg (4 kPa)
Controls the upper scale of the pressure
waveform.
The lower scale value is either
–5 mmHg (–0.7 kPa) for pressures
labeled CVP, RA, LA or 0 mmHg (0 kPa)
for other pressure labels.