Troubleshooting
!
%0 Unplugged
1), 2)
The parameter
value is replaced by
***
The pressure
transducer for
the specified
parameter is
either
unplugged or
faulty.
– During an active
pressure: Reconnect
or replace the cable.
– During an inactive
pressure: deactivate
alarms.
!
HemoPod
unplugged
2)
The parameter
value is replaced by
***
The IBP pod is
disconnected.
Check the equipment
and replace if necessary.
None
%0 zero
accepted
1)
Parameter value
Transducer
zeroing was
successful.
Informational message –
no action required.
None
%0 did not zero -
offset error
1)
Parameter value
Transducer
zeroing failed
because static
pressure was
too high or too
low.
– Keep all tubing
motionless.
– Replace the
transducer.
– Check the stopcock
and zero again.
None
Inflate balloon.
Press "Wedge" to
Start.
This message
appears in the
Wedge dialog only.
Parameter value
Action required
to start wedge
measurement.
Press
Start wedge
button to begin wedge
measurement.
None
Wedge in progress
This message
appears in the
Wedge dialog only.
Parameter value
Informational
message.
Informational message –
no action required.
1)
%0 is a placeholder for the respective IBP label (including CPP).
2)
After discharging a patient or starting the device, the alarm becomes active only after a numeric value
has been received for that parameter.
Priority
Message
Parameter field
Problem
Solution