External devices – MEDIBUS.X devices
418
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
CO
2
setup functions
See page 414 for information on how to access this
page. The tab is labelled
CO
2
when a ventilator is
connected. The tab is labelled
CO
2
/O2/ Agent
when an anesthesia workstation is connected.
Selection
Available settings
Description
Menu selections when
Ventilator
tab is selected.
CO
2
Scale
– 0 to 40 (default), 0 to 60, 0 to 80,
0 to 100 mmHg
– 0.0 to 5.0 (default), 0.0 to 8.0, 0.0 to 12.0,
0.0 to 16.0 kPa
– 0 to 5 (default), 0 to 8, 0 to 12, 0 to 16%
Determines the scale of the
displayed CO
2
waveform.
Color
Red, green, blue, yellow (default), light blue,
purple, orange, white.
Determines the color of the CO
2
parameter field and waveform.
Menu selections when
Anesthesia workstation
tab is selected.
etCO2 parameter
field
– etCO2
(default)
– etCO2/O2
Determines the appearance of
the etCO
2
parameter field.
CO
2
Scale
– 0 to 40 (default), 0 to 60, 0 to 80,
0 to 100 mmHg
– 0.0 to 5.0 (default), 0.0 to 8.0, 0.0 to 12.0,
0.0 to 16.0 kPa
– 0 to 5 (default), 0 to 8, 0 to 12, 0 to 16%
Determines the scale of the
displayed CO
2
waveform.
Color
Red, green, blue, yellow (default), light blue,
purple, orange, white.
Determines the color of the CO
2
parameter field and waveform.
O2 parameter
field
– O2
– O2/N2O
(default)
– Off
Determines the appearance of
the O
2
parameter field.
Agent parameter
field
– Agent
– Agent/ xMAC
(default)
– Agent/N2O
– Off
Determines the appearance of
the Agent parameter field.
Note: If two agents are
detected, both are displayed
automatically.