Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
363
Scio Monitoring
O
2
alarm
limits
For gas analyzers with O
2
monitoring, the
O
2
tab
displays fields on the upper left side of the window
for
etO2
and
FiO2
alarm limit
settings.
etO2
and
FiO2
allow adjustment of the upper and
lower alarm limits, use of the
Archive
feature, and
use of
Auto set.
O
2
parameter setup functions
Setup functions for O
2
parameters take place in the
O
2
tab within the
Scio
dialog.
NOTE
When a gas analyzer without O
2
monitoring is
connected, the controls for O
2
alarms are not
available in the Scio setup menu.
Selection
Available settings
Description
Parameter field
– O
2
– O2/N
2
O
Configures the O
2
display based on the
selected setting.
Scale [%]
– 0-100
– 15-35
– 20-100 (default)
– 25-45
– 35-55
– 45-65
– 55-75
– 65-85
– 75-95
– 85-105
Adjusts the size of the O
2
waveform.
Mini-trend
– etO2
(default)
– FiO2
Selects the parameter to display in the mini-
trend associated with the O
2
parameter field.