Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
463
System configuration
"All alarms off"
reminder
–
On
(default)
– Off
When the
All alarms paused
setting is set to
No timeout
(see page 459) and you select the
All alarms off
button, this feature reminds you
that alarm monitoring has been deactivated for
all alarms.
Whether or not the Cockpit is in OR
mode, the
volume of the alarm tone corresponds to the
Alarm volume
setting of 50%.
–
On
– an alarm tone sounds every
30 seconds during an alarm condition. In
addition, the message area flashes red
three times with the message
All alarms off
and the symbol
. The alarm message
field appears solid until the 30 second pass.
Then the tone sounds again and the
message and the symbol flash again.
On remote devices, the message does not
flash but appears on solid red background
with the identical message and symbol.
When cardiac bypass mode is activated,
this feature is not available.
Any changes to this setting on the Cockpit
are adopted when the M540 docks. You can
also change this setting from the M540 if it is
docked and it will update the setting on the
Cockpit.
–
Off
– No alarm tone sounds when alarm
monitoring is deactivated and an alarm
occurs.
Tone set
–
Infinity
– IEC fast
(default)
– IEC slow
Determines the type of alarm tone used (for
more information, see “Acoustic alarm
signals“on page 114).
Selection
Available settings
Description