ECG, arrhythmia, and ST segment
200
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
ECG precautions
Refer to the following sections for general
precautions:
– “Electrical safety“on page 16
– "Electrosurgery" on page 20
– “Defibrillator precautions“on page 20
WARNING
Do not select TruST leads for ECG signal
processing. If the QRS morphology of a TruST
lead differs from that of its equivalent
conventional lead, always refer to the
conventional lead.
WARNING
To prevent patient injury, always verify the
timing of the QRS synchronization pulse
before attempting cardioversion using the
Infinity MCable – Analog/Sync.
WARNING
Do not rely solely on the ECG when
monitoring seizure-prone patients. Electrical
artifacts of non-cardiac origin, such as
seizure, may prevent detection of certain
arrhythmias.
NOTE
Use of the rest ECG report is required for an ECG
signal quality that is compliant with IEC 60601-2-
25 diagnostic high frequency specifications.