![Dräger Infinity Acute Care System M540 Скачать руководство пользователя страница 476](http://html.mh-extra.com/html/drager/infinity-acute-care-system-m540/infinity-acute-care-system-m540_instructions-for-use-manual_2526773476.webp)
System configuration
476
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Biomed setup
–
network setup
The following table lists the available settings for
configuring the Infinity and the hospital network
settings in the
Infinity
and the
Hospital
pages. To
access the pages, see page 470. After making the
desired changes, select the
Apply
button to
activate them.
Selection
Available settings
Description
DHCP
–
Disabled
(default)
– Enabled
Applies to the hospital network.
When the Dynamic Host Configuration
Protocol (DHCP) is activated, the settings for
IP address
,
Subnet mask
,
Gateway
, and
Primary DNS
are pulled automatically from
the server.
IP address
User selectable
Allows you to select an IP address manually
(the
DHCP
setting has to be set to
Disabled
).
Subnet mask
User selectable
Allows you to set up a subnet mask (the
DHCP
setting has to be set to
Disabled
).
Gateway
User selectable
Allows you to set up a gateway (the
DHCP
setting has to be set to
Disabled
).
Primary DNS
User selectable
Allows you to set up the primary Domain Name
System (DNS) – set the
DHCP
setting to
Disabled
.
Offline detection
–
On
(default)
– Off
Determines if the Cockpit issues an alarm tone
and a message when it loses its connection to
the Infinity network. For more information, see
"Communicating with the Infinity network"
on page 48.