Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
511
Troubleshooting
!
External device
disconnected
An external device is no
longer communicating with
the Cockpit due to a
disconnected cable
(hardware- related).
Check the external device
connections.
An external device is no
longer communicating with
the Cockpit. When this
happens repeatedly in
rapid succession, the
Cockpit will stop monitoring
the external connection
(software-related). In this
instance all connected
devices become
disconnected.
First check the external device
connection and reconnect the
cable, if necessary. If the
connection is still not restored,
turn the Cockpit off, and then turn
it on again.
!
Not Monitored By
Central
The Cockpit is connected to
the Infinity network but is
not assigned to a central
station.
A wireless M540 is out of
range of the access point.
Admit the patient at the central
station.
Return the M540 inside the range
of the wireless access point.
Check network connectivity.
!
Offline
The Cockpit is
disconnected from the
Infinity network.
Check the network connectivity.
!!
please plug in power
supply
Loss of AC power forces
the Cockpit to run on
battery charge power for at
least five minutes before
shutting down.
Check the power source and all
connections.
!
Please plug in system
cable
The system cable was
disconnected from the
M500.
Reconnect the system cable.
!!
Power supply
overheating
The power supply is
overheating.
Unplug the power supply and
contact DrägerService
®
.
!!
Power supply H/W
failure
Faulty power supply.
Replace the power supply and
contact DrägerService
®
.
!!
Power supply low
battery
The battery charge is
< 20%. When the battery
charge falls below 10%, the
Cockpit performs a safe
shutdown.
Reconnect to AC power.
Priority
Message
Cause
Remedy