Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
479
System configuration
Biomed IT setup
The IT page consists of several setup pages for
performing the following IT-specific tasks:
– Activating or deactivating all IT tabs
– Configuring a browser
– Configuring Citrix applications
– Configuring IT applications
Activating/deactivating IT tab feature
When the web enabled tabs option is unlocked, the
Cockpit supports IT applications (options) that are
accessible via IT tabs (see “Supported IT
applications“on page 506). When an IT application
is configured and the tab is activated, the
corresponding IT tab appears along the left edge of
the screen as soon as the IT tab feature is
activated. Regardless of how many IT tabs are
configured, the top tab is always labeled
Patient
and provides access to the Cockpit main screen.
To activate or deactivate IT tab feature
1
Access the
IT setup
page (see page 470).
2
Select the
IT tabs
button.
3
Select the
On
or
Off
button next to the
IT tabs
selection.
Configuring IT tabs – browser setup
You can set up a browser as an IT tab that contains
several pre-configured Web pages. These Web
pages are accessible from a pull-down dialog under
the configured IT tab (see “Accessing an IT tab“on
page 505).
CAUTION
The Infinity Acute Care System – Monitoring Ap-
plications (IACS) does not have virus protection
software and relies therefore on the firewall of
your institution to prevent access to infected files.
While setting up IT applications to access web-
sites, evaluate each website with regard to possi-
ble virus infection.