Reports/recordings
496
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Automatic alarm recordings
When the
Alarm Waveform
feature is activated
(see page 467), timed alarm recordings are
generated automatically whenever a parameter
whose archive function is activated goes beyond
the set alarm limits.
Alarm recordings are also generated when an
arrhythmia event with an alarm classification of
high or medium occurs.
The following sections describe how to set up a
parameter or arrhythmia event to generate an
automatic alarm recording.
To activate or deactivate the archive function of
a parameter
1
Select the parameter field of the parameter
whose alarm function you wish to activate or
deactivate to access that parameter page
directly.
or
Select
Sensor parameters...
from the main
menu bar > select the desired parameter tab to
access the page.
2
Select the button next to the
Archive
setting
and select either
Store
,
Str/Rec
,
Record
to
generate a recording or
Off
to deactivate the
feature.
To assign an alarm priority to arrhythmia events
1
Select the
Alarms...
button on the main menu
bar.
2
Select the
Limits
tab (if not already selected).
3
Select the
ARR
tab along the right side to
display the
ARR
page.
4
Touch the field in the
Alarm
column of the
parameter whose alarm priority you want to
modify.
5
Select the field in the
Archive
column and
select either
Store
,
Str/Rec
,
Record
to
generate a recording or
Off
to deactivate the
feature.
6
Press the rotary knob to confirm the setting.