Reprocessing
576
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Disassembly
Observe before disassembly:
1
Switch off the device and all devices connected
to it.
2
Disconnect the mains plugs.
WARNING
Because of the risk of electric shock, never
remove the cover of any device while it is in
operation or connected to power.
WARNING
Do not immerse or rinse the device and its
peripherals. If you spill liquid on the device
(including the battery or accessories), or
accidentally immerse it in liquid, disconnect
the device from the power source and allow it
to dry completely for at least 24 to 48 hours.
Contact Dräger-authorized service personnel
regarding the continued safety of the device
and its peripherals before placing it back in
operation.
CAUTION
To avoid damaging the device, do not use sharp
tools or abrasives. Never immerse electrical
connectors in water or other liquids.