Technical data
586
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Overview
This chapter contains the technical data for the
following devices of the Infinity Acute Care System
– Monitoring applications:
– PS250 power supply
– P2500 power supply
– Minimal technical requirements for the
secondary display
– Infinity MCable – Nurse call
For technical data of the Infinity C500/C700 refer to
instructions for use
Infinity Medical Cockpit
.
For the following information, refer to the
instructions for use
Infinity Acute Care System –
Infinity M540
:
– Infinity M500 docking station
– MPod and MCable devices that connect directly
to the M540
– Specifications such as measuring ranges of
individual parameters
The IACS is intended to be connected to one
patient at a time.
Device combinations
This device can be operated in combination with
other Dräger devices or with devices from other
manufacturers. Observe the accompanying
documents of the individual devices.
If a device combination is not approved by Dräger,
the safety and the functional integrity of the
individual devices can be compromised. The
operating organization must ensure that the device
combination complies with the applicable editions
of the relevant standards for medical devices.
Device combinations approved by Dräger meet the
requirements of the following standards:
– IEC 60601-1, 3rd edition (general requirements
for safety, device combinations, software-
controlled functions)
– IEC 60601-1-2 (electromagnetic
compatibility)
– IEC 60601-1-8 (alarm systems)
Or:
– IEC 60601-1, 2nd edition (general
requirements for safety)
– IEC 60601-1-1 (device combinations)
– IEC 60601-1-2 (electromagnetic
compatibility)
– IEC 60601-1-4 (software-controlled
functions)
– IEC 60601-1-8 (alarm systems)
If a device combination is not approved by Dräger,
functional integrity of the devices can be
compromised.
The operating organization must ensure that the
device combination meets the applicable
standards.
Strictly observe instructions for use and assembly
instructions of all connected devices.
CAUTION
Combinations of Dräger devices and third-party
devices that are not approved by Dräger may
adversely affect operation of those devices and
may put the patient at greater risk of injury.