System configuration
472
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Biomed setup
–
units of measure
The following table lists the available settings of the
Units of measure
page where you can configure
the units for all parameter groups. To access this
page, see page 470. Select the
Apply
button after
making your selection.
CAUTION
Before you connect the M540 to a different Cock-
pit, make sure that the units of measure align be-
tween the two devices. Differing units of measure
could result in loss of data or a patient discharge.
Selection
Available settings
Description
Pressure
–
mmHg
(default)
– kPa
Assigns the selected unit of measurement to
the parameter. Whenever you change a unit of
measurement, the Cockpit discharges the
patient.
CO2 –
mmHg
(default)
–
kPa
– %
Assigns the selected unit of measurement to
the parameter. Whenever you change a unit of
measurement, the Cockpit discharges the
patient.
Applies to Mainstream and Microstream
(sidestream).
Temperature
–
°C
(Celsius) default
–
°F
(Fahrenheit)
Assigns the selected unit of measurement to
the parameter. Whenever you change a unit of
measurement, the Cockpit discharges the
patient.
ST
–
mm
(default)
– mV
SpHb
(only Masimo
rainbow SET)
– g/dL
– mmol/L
Agent
–
kPa
(default)
– %
Weight
–
kg
(default)
–
lb
(adult, pediatric)
–
oz
,
g
(neonate)
Height
–
cm
(default)
– in