![Dräger Infinity Acute Care System M540 Скачать руководство пользователя страница 164](http://html.mh-extra.com/html/drager/infinity-acute-care-system-m540/infinity-acute-care-system-m540_instructions-for-use-manual_2526773164.webp)
Trends/data dialogs
164
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Special characters and symbols
In addition to parameters, certain conditions, such
as disconnected leads, artifact, and so on, are also
identified on graphical and tabular trends.
Graphical trends
A graphical trend plots the behavior of parameters
over time. Graphical trends are continuously
updated, with the most recent data appearing on
the right side of the screen. The
Trends/Data
dialog consists of the following graphical trend
pages:
–
Graph
page
–
Graph vitals
page
–
Ventilation / Anesthesia
page
All graphical trend pages look almost identical. In
the
Graph
and the
Ventilation / Anesthesia
pages
you can change the parameter content.
However, the
Graph vitals
is a pre-configured
display consisting of the following set of commonly
trended parameters which are displayed in four
graphical windows:
To access the graphical trend pages
1
Select the
Trends/ Data...
button on the main
menu bar.
2
Select the
Trends
tab (if not already selected).
3
Select one of the following graphical trend
pages:
–
Graph
tab to view general trends
–
Graph vitals
tab to view a set of
pre-configured parameters
–
Ventilation / Anesthesia
tab to view
ventilation/anesthesia-related parameters
–
BIS
tab to view BIS trends
Event
Character/symbol
Asystole
ASY
Ventricular fibrillation
VF
Apnea
Apnea
No parameter values are
available
* * *
Out of range value
+++ (high) - - - (low)
Relearning
LEARN
Interruption in power or
monitor is placed into
standby
No values
Window 1 (top)
HR, SpO2, and RRi
Window 2
NIBP
Window 3
Ta, Tb
Window 4
CO
2