Invasive blood pressure (IBP)
302
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Overview of invasive blood pressure monitoring
The M540 acquires, processes, and displays
invasive blood pressure (IBP) signals and relays
the data to the Cockpit. Several pods are available
for monitoring invasive blood pressure. Monitoring
more than two pressures simultaneously requires
the Multi-IBP option.
Invasive blood pressure measurements are for
adult, pediatric, and neonatal patients.
Refer to the Instructions for use
Infinity Acute Care
System – Infinity M540
for a detailed description of
the M540 invasive blood pressure functions.
The invasive blood pressure monitoring functions
are configurable in the parameter-specific setup
page (see "Invasive blood pressure parameter
setup functions" on page 319).
Before performing any monitoring functions, refer
to the section "For your safety and that of your
patients" on page 13. Parameter-specific error
messages are listed on page 539.
Supported parameters
See page 310 for available invasive blood pressure
labels.
– Systolic pressures: GP1 S to GP4 S, ART S,
PA S, LV S, RV S
– Diastolic pressures: GP1 D to GP4 D, ART D,
PA D, LV D, RV D
– Mean pressures: GP1 M to GP4 M, ART M,
PA M, LV M, RV M
– Additional pressures: ICP, CVP, LA, RA
– If both ART and ICP are connected, the
algorithm computes the difference between ICP
and mean ART and reports it as CPP.
Hemodynamic pods
Invasive blood pressure signals originate from the
following hemodynamic pods:
– Hemo4
– Hemo2 pod
– Infinity MPod – QuadHemo (MPod –
QuadHemo)
– Infinity MCable – Dual Hemo (Dual Hemo
MCable)
Hemo4 pod
This pod measures up to four pressures, cardiac
output, and temperature.
A
Key
for zeroing all pressures
simultaneously (see page 312)
B
Key for starting a cardiac output measurement
C
Key for starting wedge pressure measurements
D
Pressure label windows
E
Transducer slots
338
NOTE
The ports for temperature and cardiac output are
located on the side of the hemodynamic pod.
A B
C
D
E
E
E
E