Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
279
SpO2 and pulse rate with Nellcor OxiMax MCable
Response mode
–
Normal
(default) – up to
90% change within 5 to
7 seconds
–
Fast
– up to 90% change
within 2 to 4 seconds
Establishes the frequency the oximeter uses
to calculate, record, and display SpO2
saturation levels:
– Fast mode responds to changes in blood
oxygen saturation levels in 2 to 4 seconds
when calculating%SpO2.
– Normal mode responds to changes in
blood oxygen saturation in 5 to 7 seconds
when calculating%SpO2.
SatSeconds alarm
Off
(default), 10, 25, 50, 100
SatSeconds
NOTE:
When
SatSeconds
is
set to any value other than
Off
,
the
Desat.
alarm status is set to
Off
.
This selection does the following:
– Analyzes desaturation events by
multiplying their duration (seconds) by the
number of percentage points the patient
exceeds the alarm limit.
– Eliminates nuisance alarms caused by
brief and numerous violations of lower and
upper alarm limits.
– Overrides the alarm validation setting
(see page 459) and the SpO2 high priority
desaturation alarm.
Color
Red, green, blue, yellow, light
blue, purple, orange, white
(default).
Determines the color of the waveforms and
parameter labels and values.
Selection
Available settings
Description
NOTE
The password-protected alarm setting
SpO2 check sensor
provides additional SpO2 alarm
configuration. For more detailed information see page 117.