Maintenance
568
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Overview
This chapter describes the maintenance measures
required to maintain the functional integrity of the
medical device. Maintenance measures must be
performed by the responsible personnel.
WARNING
Risk of infection.
Users and service personnel can become
infected with pathogens.
Disinfect and clean the device or the
components before any maintenance
measures and also before returning the
medical device for repair.
WARNING
Risk of electric shock.
Current-carrying components are located
under the cover.
– Do not remove the cover.
– Maintenance measures must be performed
by the responsible personnel. Dräger
recommends DrägerService to perform
these measures.
WARNING
If the device is mechanically damaged, or if it
is not working properly, do not use it. Contact
your hospital’s service personnel.
CAUTION
This device must be inspected and serviced at
regular intervals. A record must be kept on this
preventive maintenance. We recommend obtain-
ing a service contract with DrägerService through
your vendor. For repairs we recommend that you
contact DrägerService.
CAUTION
When servicing devices from Dräger, always use
spare parts that are qualified to Dräger standards.
Dräger cannot warrant or endorse the safe
performance of third-party spare parts for use with
the devices.
CAUTION
If you spill liquid on the equipment, battery or
accessories or immerse these components in
liquid, allow them to dry completely for at least 24
hours to 48 hours. Contact your hospital’s service
personnel to test any such component is fully
operational before putting it back in clinical use.
NOTE
Only perform maintenance measures when no
patient is connected to the device.
WARNING
Any modification of this device or any use dif-
ferent from the one specified in these instruc-
tions for use may cause interference with oth-
er equipment. It may also result in injury to the
patient or the user, including electric shock,
burns or death.