![Dräger Infinity Acute Care System M540 Скачать руководство пользователя страница 558](http://html.mh-extra.com/html/drager/infinity-acute-care-system-m540/infinity-acute-care-system-m540_instructions-for-use-manual_2526773558.webp)
Troubleshooting
558
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Low
CO
2
reduced
accuracy
Accuracy of the CO
2
sensor
cannot currently be
guaranteed.
– Ensure clean ambi-
ent air
– Check water trap and
sample line.
– Change the water
trap or sample line if
necessary.
– Wait for automatic ze-
roing
– Power cycle the gas
analyzer.
– Call DrägerService
®
.
Low
%0 sensor failure
The CO
2
sensor in patient gas
measurement module has
failed due to:
– Sample line occlusion.
– Electrical disturbance.
– Internal failure.
– Check sample line.
– Remove radiating
devices (e.g.,
telephone).
– Use alternative CO
2
measurement
system.
– Call DrägerService
®
.
Medium
%0 out of range high
CO
2
concentration has
exceeded the Scio upper limit
of the measurement range.
Check vaporizer, fresh-
gas settings and ventila-
tion.
Low
etAgent < #
Note
This alarm only occurs
for the primary agent.
– Expiratory anesthetic gas
concentration has fallen
below the lower alarm limit
for more than 15 seconds.
– Soda lime is dried out.
– Check vaporizer and
fresh-gas settings.
– Check breathing
system for large
leaks.
– Exchange soda lime.
Medium
etAgent > #
Note
This alarm only occurs
for the primary agent.
Expiratory anesthetic gas
concentration has exceeded
the upper alarm limit for more
than 15 seconds.
Check vaporizer and
fresh-gas settings
Medium
etCO
2
< #
Expiratory CO
2
has fallen be-
low the limit for more than 15
seconds.
Check ventilation.
Alarm Priority Alarm
Cause
Remedy