ECG, arrhythmia, and ST segment
232
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
ST reference
You can save ST reference complexes as
reference points for future ST deviation
measurement comparisons. The first time you
relearn QRS complexes, the current ST data are
saved as a reference data. The original ST
reference data are updated each time you save ST
references.
Saving ST reference points
You can save the ST reference from the general ST
complexes page (see page 229) and the single ST
complex page (see page 229). Saving a reference
point in either screen, saves all currently displayed
ST complexes as reference points.
To save ST reference points
Select the heart rate parameter field to select
the
ECG
page directly.
or
1
Select
Sensor parameters...
from the main
menu bar.
2
Select the
ECG
tab to access the
ECG
page.
3
Select the
ST complex
tab to display the
general
ST complex
page.
4
Select an individual ST panel to zoom in on a
single ST complex.
5
Select the
ISO
and
ST
buttons at the bottom of
the screen and use the rotary knob to dial the
desired values. Click on the rotary knob to
accept the new values.
6
Select the
Save reference
button (in either ST
complex page).