Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
149
Alarms
Alarm ranges and defaults
Parameter Alarm
default
status
Alarm limit range
Upper limit
defaults
Lower limit
defaults
Archive
default
setting
HR adult
Increment:
5 bpm
On
Upper: 25 to 300 bpm
Lower: 20 to 295 bpm
120 (adult)
150 (pediatric)
170 (neonate)
45 (adult)
50 (pediatric)
80 (neonate)
Str/Rec
(adult/
pediatric)
Off
(neonate)
STVM/STCVM
Increment:
0.1 mm or
0.01 mV
Upper: 0.1 to 45.0 mm
0.01 to 4.50 mV
Lower: 0.0 to 44.9 mm
0.00 to 4.49 mV
1.0 mm (0.1 mV) 0.0 mm
(0 mV)
Off
ST
Increment:
0.1 mm or
0.01 mV
Upper: –14.9 to +15.0 mm
–1.49 to +1.50 mV
Lower: –15.0 to +14.9 mm
–1.50 to +1.49 mV
1.0 mm
(0.1 mV)
–1.0 mm
(–0.1 mV)
Off
RRi (adult)
Increment: 1
Upper: 6 to 100
Lower: 5 to 99 (adult)
30
5
Off
RRi (pediatric,
neonate)
Increment: 1
Upper: 6 to 145
Lower: 5 to 144
80
20
Off
PLS
Increment of 5
Upper: 35 to 235
Lower: 30 to 230
120 (adult)
150 (pediatric)
180 (neonate)
45 (adult)
50 (pediatric)
80 (neonate)
Off