Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
103
Getting started
Admitting a patient using
Get HIS
You can populate the
Demographics
page
automatically, by pulling the demographic data of a
patient from the network. Prerequisite for this
network data transfer is the Infinity gateway with an
interface to the hospital Admit, Discharge, Transfer
(ADT) system. The Hospital Information System
(HIS) searches the database for the demographic
data of the patient by using the patient ID.
To admit a patient via
Get HIS
Touch the left most field on the header bar to ac-
cess the
Demographics
page directly.
or
1
Select
Start/Standby...
on the main menu bar.
2
Select the
Demographics
tab (if not already
selected).
3
Enter the patient ID – use the symbol (C) next
to the
Patient ID
field to activate a keyboard for
entering the ID number (up to 13 alphanumeric
characters).
4
Select the
Get HIS
button (E) in the
Demographics
page (see page 101). The
Get
HIS
button appears grayed out and is not
selectable when the HIS is not available or
when the Cockpit is not connected to it.
Discharging a patient
You can discharge a patient from the Cockpit or
from the M540. Discharging a patient from either
device causes a discharge at the other device.
Refer to the instructions for use
Infinity Acute Care
System – Infinity M540
, for detailed information on
how to discharge a patient from the M540.
Discharging a patient has the following effect on the
Cockpit:
– All demographic data are removed from the
screen
– All trend and event data are deleted
– Any active recordings are canceled
– The profile with defined patient settings is re-
stored
– The message
Touch Screen to initiate moni-
toring
appears