Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
529
Troubleshooting
!
(Masimo
rainbow
SET only)
SpO2 cable failure
Cable failure
Parameter values
are replaced by ***
The Masimo rainbow
SET intermediate
cable is faulty.
Replace the
intermediate cable.
!
(Nellcor
OxiMax
MCable
only)
SpO2 check
sensor
2)
Check sensor
Parameter values
are replaced by ***
The SpO2 sensor is
disconnected.
– Make sure the
SpO2 sensor is
attached
properly to the
patient.
– Check all cable
connections.
None
(Masimo
rainbow
SET only)
%0 sensor
calibrating
1)
Sensor
calibrating
Parameter values
are replaced by ***
The sensor is being
checked for functional
integrity.
Wait until message
disappears.
This message
appears right
before the
message
SpO2
searching
.
None
(Any
SpO2
MCable)
SpO2 desaturation
Parameter value
SpO
2
value below
Desat. limit.
– Check the
patient and
treat if
necessary.
– Change the
alarm limits.
Priority
Message
Parameter field
Problem
Solution