![Dräger Infinity Acute Care System M540 Скачать руководство пользователя страница 229](http://html.mh-extra.com/html/drager/infinity-acute-care-system-m540/infinity-acute-care-system-m540_instructions-for-use-manual_2526773229.webp)
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
229
ECG, arrhythmia, and ST segment
Reviewing ST complexes
You can view all ST complexes or zoom in on a
single complex.
The following functions are available in either view:
– Changing the isoelectric point
– Changing the ST measuring point
– Relearning the QRS morphology
– Saving a reference complex
– Requesting an ST report
In all trends, a solid white vertical line on the ST
graphical trends marks changes in ST measuring
points along with a time stamp.
Reviewing all ST
complexes
The following diagram shows the
ST complex
page. The number of displayed ST complexes
depends on the connected lead set.
A
ECG
tab
B
ST complex
tab
C
Save reference
button
D
button for generating an ST report
E
ST
button
F
Reference
on/off
button
G
ISO
button
H
Relearn
button
I
ST panels for each monitored ST lead
(including waveform scale, ST measuring point,
isoelectric point, and reference point)
J
ST label (unique for each ST lead)
To access ST complexes
Select the heart rate parameter field to select
the
ECG
page directly.
or
1
Select
Sensor parameters...
from the main
menu bar.
2
Select the
ECG
tab to access the
ECG
page.
3
Select the
ECG
tab (if not already selected).
4
Select the
ST complex
tab (if not already
selected).
035
A
B
I
J
C
D
E
F
G