External devices – MEDIBUS.X devices
414
Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
2
Press any button to activate a list of available
parameters. You can also select
None
to
remove any parameter from being displayed in
that space. The parameter selection
MV auto
,
RR auto
and
VT auto
are unique because the
associated parameter set varies from device to
device. When the
Vent parameter display
setting
Auto
is selected, the Cockpit pulls the
available parameter from the source device
according to the following priority list:
–
MV auto
:
MVe, MV,MVi
–
RR auto
: RR, RRf, RRp
–
VT auto
: VTe, VTACH,VTi
3
Select the desired parameter in the list to assign
it to the button and to the
Show all
page.
4
Repeat steps 2 and 3 until the desired configu-
ration for the
Show all
page is completed.
Accessing parameter setup functions
You can configure the display of ventilation and
anesthesia parameters in the following tabs located
under the
Ventilator
or the
Anesthesia
workstation
tabs.
– Paw
– Vent
–
CO
2
(when a ventilator is connected) or
CO
2
/O2/ Agent (when anesthesia workstation is
connected)
1
Select
Sensor parameters...
from the main
menu bar.
2
Select the >
Ventilator
or the
Anesthesia
workstation
tab.
If you cannot see the tab, select the following
two symbols located in the upper right corner of
the dialog:
symbol and the display filter
button.
3
Select either the
Paw
,
Vent
or
CO
2
or
CO
2
/O2/
Agent
tabs to access the respective setup
pages.
NOTE
Depending on which devices are connected to the
Cockpit, two
Ventilator
tabs may be displayed.
The MEDIBUS.X
Ventilator
tab is the one that
has the message
Medibus.X ventilator devices
displayed as a header on the various pages.
NOTE
Most setup functions described in the following ta-
bles are available under the
Ventilator
and the
Anesthesia workstation
tabs. Exception where
a setup function is only available under one tab
are noted.