18-461
18-461
18.7. Force sensor calibration
18.7.1. Outline
Force sensor calibration is the function that calculates the weight and center of gravity position of the robot hand
attached to a force sensor.
The data calculated by force sensor calibration (robot hand weight and center of gravity position) is needed
during force offset cancellation. Execute the force sensor calibration before the execution of gravity offset
cancellation.
The table below shows the robot controller versions and robot models.
Table 18-14 Supported robot controllers and model
No.
Robot model
CR800
CR750/700
CRn-500
1
Vertical 6-axis robot
All versions
(RV-FR series models)
Version R6h/S6h
or later
(RV-F series models)
not used
2
Vertical 5-axis robot
not used
not used
3
Horizontal 4-axis robot
not used
not used
18.7.2. Starting
From the project tree, double-click [Tool] -> [Force sensor] -> [Force sensor calibration]
Figure 18-69 Startup screen of force sensor calibration
This function can be used when connected with a machine supporting force sensor function. For details on the
setup method, refer to the user’s manual of the force sensor function (BFP-A8940).
Содержание 3F-14C-WINE
Страница 84: ...8 84 8 84 Figure 8 21 Parameter transfer procedure 1 2 2 ...
Страница 393: ...18 393 18 393 Figure 18 1 Starting the Oscillograph ...
Страница 413: ...18 413 18 413 Figure 18 24 Output to Robot program Selection ...
Страница 464: ...18 464 18 464 Figure 18 72 Starting the Tool automatic calculation ...
Страница 545: ...21 545 21 545 Figure 21 55 Hide display of user mechanism ...
Страница 624: ...24 624 24 624 Figure 24 4 Document output example ...