4-224
Operation
B337EN4_8
14.11Carrying out a leakage test
Select the "DIAGNOSTICS - Laser - Leakage test" activity field.
Use the focus key to select the "Start pressure" display field.
7
4
1
2
3
5
6
8
9
0
Use the keyboard to enter the start pressure.
Use the focus key to select the "Measuring time" display field.
7
4
1
2
3
5
6
8
9
0
Use the keyboard to enter the start pressure.
The default measuring time is 2 minutes.
Start
leak test
Press the Start leak test softkey.
Wait until the message "TLF Laser: Leakage test" ended appears.
Prerequisite
•
Laser off.
•
The resonator is in idle mode.