![Rohde & Schwarz SMBV100A Operating Manual Download Page 818](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735818.webp)
Remote Control Commands
R&S
®
SMBV100A
803
Operating Manual 1176.8016.02 ─ 17
Parameters:
<Mode>
AUTO | EXTernal | EGATe
AUTO
The pulse modulation is generated continuously.
EXTernal
The pulse modulation is triggered by an external trigger event.
The trigger signal is supplied via the PULSE EXT connector.
EGATe
The pulse modulation is gated by an external gate signal. The
signal is supplied via the PULSE EXT connector.
*RST:
AUTO
Example:
PULM:TRIG:MODE EXT
selects triggering by an external trigger event.
Options:
R&S SMBV-K23 (Pulse Generator)
Manual operation:
See
" Trigger Mode - Pulse Generator "
[:SOURce<hw>]:PULM:WIDTh
<Width>
Sets the width of the generated pulse. The width determines the pulse length. The
pulse width must be at least 20ns less than the set pulse period.
Parameters:
<Width>
float
Range:
10 ns to 100 s
Increment: 10 ns
*RST:
2 us
Example:
PULM:WIDT 33 us
sets a width of 33 us for the pulse.
Options:
R&S
SMBV-K23 (Pulse Generator)
Manual operation:
See
" Pulse Width - Pulse Generator "
7.16.19 SOURce:ROSCillator Subsystem
This subsystem contains the commands for setting the external and internal reference
frequency.
The settings of the reference oscillator are not affected by an instrument reset (
on page 507). They are only reset to factory state by the factory-preset (
[:SOURce]:ROSCillator:EXTernal:FREQuency
..................................................................804
[:SOURce]:ROSCillator:EXTernal:RFOFf[:STATe]
............................................................. 804
[:SOURce]:ROSCillator:EXTernal:SBANdwidth
................................................................. 804
SOURce Subsystem