Remote Control Commands
R&S
®
SMBV100A
754
Operating Manual 1176.8016.02 ─ 17
●
start a new signa lgeneration
●
set a new output level
Return values:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
SOUR:IQ:OUTP:DIG:OFL:HOLD:STAT?
Response: "0" (no overflow)
Usage:
Query only
Manual operation:
See
"Overflow Hold - Digital I/Q Out"
[:SOURce]:IQ:OUTPut:DIGital:OFLow:STATe?
Queries the current overflow state of the output signal and indicates that the I/Q output
signal is currently clipped.
Return values:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
IQ:OUTP:DIG:OFL:STAT?
queries the overflow state.
Response: "0" (no overflow).
Usage:
Query only
Manual operation:
See
[:SOURce<hw>]:IQ:OUTPut:DIGital:PON
<Pon>
Selects the state which the digital I/Q Output is to assume after the instrument is
switched on.
Parameters:
<Pon>
OFF | UNCHanged
*RST:
OFF
Example:
SOURce:IQ:OUTPut:DIGital:PON OFF
deactivates the output when the instrument is switched on.
Manual operation:
See
"Power-On State - Digital I/Q Out"
[:SOURce<hw>]:IQ:OUTPut:DIGital:POWer:LEVel
<Level>
Sets the rms level of the digital output signal.
SOURce Subsystem