Operating Manual 1176.8016.02 ─ 17
"Off"
The signal output begins after elapsing of the processing time and
starts with sample 0, i.e. the complete signal is outputted.
This mode is recommended for triggering of short signal sequences
with signal duration comparable with the processing time of the
instrument.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TRIGger[:EXTernal]:SYNChronize:OUTPut
External Delay Unit
Determine whatever the trigger delay is expressed in samples or directly defined as a
time period (seconds).
The parameter displays the delay converted in time.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TRIGger:DELay:UNIT
Trigger Delay
Sets the trigger signal delay in samples on external triggering or on internal triggering
via the second path. This enables the R&S
SMBV to be synchronized with the device
under test or other external devices.
The parameter
displays the delay converted in time.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TRIGger[:EXTernal<1|2>]:DELay
[:SOURce<hw>]:BB:ARBitrary:TRIGger[:EXTernal<1|2>]:TDELay
Actual Trigger Delay
Displays the time (in seconds) an external trigger event or a trigger event form the
other path is delayed by.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TRIGger[:EXTernal<1|2>]:RDELay?