![Rohde & Schwarz SMBV100A Operating Manual Download Page 568](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735568.webp)
Remote Control Commands
R&S
®
SMBV100A
553
Operating Manual 1176.8016.02 ─ 17
Usage:
Event
:DISPlay:DIALog:ID?
Returns the dialog identifiers of the open dialogs in a string separated by blanks.
Return values:
<DialogIdList>
string
Example:
DISP:DIAL:ID?
Response:
"<dialog ID(1)> <dialog ID(2)> ...
<dialog ID(n)>"
Returns the dialog identifiers of all opened dialogs.
Usage:
Query only
:DISPlay:DIALog:OPEN
<DialogId>
Opens the specified dialog. To determine the dialog identifier, use command
.
Setting parameters:
<DialogId>
string
Example:
DISP:DIAL:OPEN "<dialog ID>"
Opens the dialog, determined with the "<dialog ID>".
Usage:
Setting only
:DISPlay:PSAVe:HOLDoff
<HoldoffTimeMin>
Sets the waiting time for the screen-save mode of the display.
Parameters:
<HoldoffTimeMin>
integer
Range:
1 to 60
*RST:
n.a. (factory preset: 10)
Default unit: minute
Example:
DISP:PSAV:HOLD 8
Sets the timeout of the screen saver to 8 minutes.
Manual operation:
See
:DISPlay:PSAVe[:STATe]
<State>
Activates the screen-save mode of the display.
If activated, the display including backlight is switched off after the wait time elapses
and if no entries via front panel, external mouse or external keyboard are made. To set
the wait time, use command
.
DISPlay Subsystem