Remote Control Commands
R&S
®
SMBV100A
741
Operating Manual 1176.8016.02 ─ 17
f
STARt
> f
STOP
is permitted.
f
STOP
= (f
CENTer
+ f
SPAN
/2).
Note:
A defined offset affects the sweep range and consequently all correlating param-
eters. The set frequencies are only absolute values, if the Offset = 0. Offset
≠
0 shifts
the frequencies according to the offset value.
f
STARt
* f
MULTiplier
+ f
OFFSet
... f
STOP
* f
MULTiplier
+ f
OFFSet
Parameters:
<Stop>
float
Range:
full frequency range
Increment: see the data sheet: RF characteristics > Resolution
of setting
*RST:
500 MHz
Example:
FREQ:STOP 2 GHz
sets the stop frequency for the frequency sweep to 2 GHz.
FREQ:STAR 1 MHz
sets the start frequency for the frequency sweep to 1 MHz.
Manual operation:
See
" Stop Freq - Frequency Sweep "
[:SOURce<hw>]:FREQuency:STEP[:INCRement]
<Increment>
Sets the step width for
To adjust the frequency step by step with this step size, use the
FREQ:UP
and
FREQ:DOWN
commands.
Note:
This value also applies to the step width of the rotary knob of the instrument and
increases or decreases the frequency accordingly, when you work in user-defined step
mode.
Parameters:
<Increment>
float
Range:
full frequency range
Increment: see the data sheet: RF characteristics > Resolution
of setting
*RST:
1E6
Example:
FREQ:STEP 50 kHz
sets the step width for the frequency setting to 50 kHz.
Manual operation:
See
SOURce Subsystem