Remote Control Commands
R&S
®
SMBV100A
752
Operating Manual 1176.8016.02 ─ 17
Parameters:
<Mode>
FIXed | VARiable | VATTenuated
FIXed
The settings are fixed and cannot be changed.
VARiable
The settings can be changed (command
VATTenuated
Enables a precise digital attenuation of the signal. The level of
the baseband and the RF signal is attenuated by the value set
with the command
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:
*RST:
FIXed
Manual operation:
See
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:POWer:DATTenuation
<DAttenuation>
The level of the baseband and the RF signal is attenuated by this value.
Parameters:
<DAttenuation>
float
Range:
-3.522 to 80
Increment: 1E-3
*RST:
0
Example:
SOUR:IQ:OUTP:ANAL:MODE VATT
SOUR:IQ:OUTP:ANAL:POW:DATT 10dB
The level of the baseband and RF signal is attenuated with
10dB.
Manual operation:
See
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:TYPE
<Type>
Selects an input type.
Parameters:
<Type>
SINGle | DIFFerential
SINGle
The I/Q-signal components are output single-ended.
DIFFerential
The I/Q-signal components are output differential.
*RST:
SINGle
Example:
IQ:OUTP:ANAL:TYPE DIFF
the I/Q-signal components are output differential.
Manual operation:
See
SOURce Subsystem