![Rohde & Schwarz SMBV100A Operating Manual Download Page 289](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735289.webp)
Instrument Function
R&S
®
SMBV100A
274
Operating Manual 1176.8016.02 ─ 17
"Ext Gated"
The pulse generator signal is gated by an external gate signal.
Example: Generation of pulse signals using trigger mode Exter-
nal Gated (Single Pulse)
The measurement is performed using a 6-dB-attenuator.
Parameter
Value
Trigger Mode
Extern Gated
Double Pulse State
Off
External Trigger Input Slope
Positive
Pulse Delay
100 ns
Pulse Width
100 ns
Pulse Period
300 ns
Channel 1 = Indicates the external gate signal. The signal is active when it is high
(positive).
Channel 2 = Indicates the sync signal. The sync signal starts after a trigger delay of
typically 50 ns (see specifications). It is repeated after the set pulse period
of 300 ns as long as the gate signal is active.
Channel 3 = Indicates the pulse signal. The first pulse starts after the pulse delay of
100 ns. The second pulse starts after the set pulse period.
Remote command:
[:SOURce<hw>]:PULM:TRIGger:MODE
External Trigger Input Slope - Pulse Generator
(External Trigger only)
Sets the polarity of the active slope of an applied trigger signal.
"Positive"
The pulse generator is triggered on the positive slope of the external
trigger signal.
"Negative"
The pulse generator is triggered on the negative slope of the external
trigger signal.
Remote command:
[:SOURce<hw>]:PULM:TRIGger:EXTernal:SLOPe
RF Block