Remote Control Commands
R&S
®
SMBV100A
582
Operating Manual 1176.8016.02 ─ 17
Usage:
Query only
:SENSe<ch>[:POWer]:DISPlay:PERManent:STATe
<State>
The command switches on and off the permanent indication of the power measure-
ment result in the upper right corner of the block diagram. For each sensor, the type of
sensor, the connector, the measurement source and - if set - the offset is indicated.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
ON
Example:
SENS1:POW:DISP:PERM:STAT ON
the permanent viewer is switched on.
Manual operation:
See
:SENSe<ch>[:POWer]:DISPlay:PERManent:PRIority
<Priority>
The command selects which power measurement result (average or peak power) is
indicated when permanent display is active.
Parameters:
<Priority>
AVERage | PEAK
*RST:
PEAK
Example:
SENS1:DISP:PERM:STAT ON
the permanent viewer is switched on.
SENS1:DISP:PERM:PRI AVER
the measured average power is indicated.
Manual operation:
See
:SENSe<ch>[:POWer]:FILTer:LENGth:AUTO?
The command queries the current filter length for auto filter mode
(
:SENSe<[1]...3>:POWer:FILTer:TYPE AUTO
)
Return values:
<Auto>
float
Range:
1 to 65536
Example:
SENS1:FILT:TYPE AUTO
selects auto filter mode for the power sensor connected to the
SENSOR connector.
SENS1:FILT:LENG:AUTO?
queries the automatically set filter length.
Response: 1024
Usage:
Query only
Manual operation:
See
SENSe, READ and INITiate Subsystems