![Rohde & Schwarz SMBV100A Operating Manual Download Page 770](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735770.webp)
Remote Control Commands
R&S
®
SMBV100A
755
Operating Manual 1176.8016.02 ─ 17
Parameters:
<Level>
float
Range:
-90 to 0
Increment: 0.01
*RST:
0
Default unit: dBFs
Example:
IQ:OUTP:DIG:POW:LEV -10
sets the output signal level to -10 dBFS.
Manual operation:
See
[:SOURce<hw>]:IQ:OUTPut:DIGital:POWer:PEP
<Pep>
Sets the peak envelope power of the digital output signal (PEP).
Parameters:
<Pep>
float
Range:
-60 to 0
Increment: 0.01
*RST:
0
Default unit: dBFS
Example:
IQ:OUTP:DIG:POW:PEP -10
Sets the signal output level to -10 dBFS.
Manual operation:
See
"Peak Level - Digital I/Q Out"
[:SOURce<hw>]:IQ:OUTPut:DIGital:POWer:STEP[:INCRement]
<Increment>
Sets the step width for the frequency setting if the frequency values
UP/DOWN
are used
and variation mode
SOUR:IQ:OUTP:DIG:POW:STEP:MODE USER
is selected. The
command is linked to "Variation Step" for manual operation, i.e. the command also
sets the step width of the rotary knob for "Variation Active ON".
Parameters:
<Increment>
float
Range:
0 to 100
Increment: 0.01
*RST:
1
Default unit: dB
Example:
SOUR:IQ:OUTP:ANAL:POW:STEP:INCR 50 mV
sets the step width for the level setting to 50 mV.
Manual operation:
See
"Variation Step - Digital I/Q Out"
SOURce Subsystem