![Rohde & Schwarz SMBV100A Operating Manual Download Page 414](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735414.webp)
Instrument Function
R&S
®
SMBV100A
399
Operating Manual 1176.8016.02 ─ 17
The subsequent trigger event after the last segment causes the
first segment to be output again.
"Next Segment
Seamless"
The segment selected under "Next Segment" is output.
This mode is only available if all segments have the same sample
rate.
●
In the case of "Internal Trigger = Auto", output starts at once and
the segment is generated continuously.
If the segment is changed in "Next Segment", the new segment is
output seamlessly after the output of the current segment is com-
plete.
●
In the case of "internal Trigger = Armed_Auto", output starts after
the first trigger event. The segment is then generated continu-
ously.
If the segment is changed in"Next Segment", signal output is
stopped and the new segment is not output until a trigger occurs.
●
In the case of "External Trigger = Auto", output starts at once and
the segment is generated continuously.
Each trigger event (re-)starts the segment; each next segment
trigger event switches over to outputting the subsequent segment
once the output of the current segment has been completed.
If the segment is changed in "Next Segment", the new segment is
output seamlessly after the output of the current segment is com-
plete.
●
In the case of "External Trigger = Armed_Auto", output starts after
the first trigger event. The segment is then generated continu-
ously.
Each trigger event (re-)starts the segment; each next segment
trigger event switches over to outputting the subsequent segment
once the output of the current segment has been completed.
If the segment is changed in "Next Segment", signal output is
stopped and the new segment is not output until a trigger occurs.
The remaining trigger modes ("Retrigger" and "Armed_Retrigger") are
not available.
"Sequencer"
The waveform files are processed according to the order and repeti-
tion cycles defined in the special sequencing list file (
*.wvs
)
assigned to the multi segment waveform file.
This “play list” is defined in the
Multi Segment Waveform Sequencing
dialog.
Note:
This feature is disabled in case the segments in the multi seg-
ment waveform file have different clock rates.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TRIGger:SMODe
Next Segment Source
Selects the next segment source, i.e. determines whether the next segment is
switched upon receiving of an external trigger signal or upon executing of software
next segment trigger .
Baseband Signal - Baseband Block